Model Number 72200616S
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/1901
Event Type  Malfunction  
Event Description

It was reported that doctor noticed the shaver hand piece become hot to the touch toward the end of the case. No patient injury was reported.

 
Manufacturer Narrative

The device was received for evaluation. There was no relationship found between the returned device and the reported incident. A visual inspection was performed on the product and no issue was observed. Complaint of overheating could not be reproduced. Product passed functional and high speed testing (100-10,000 rpms) for 5 minutes each in forward, reverse and oscillate directions. Unit passed functional tests on the dyonics power, dii and dii eip test control units with and without footswitch. Current draw was average for this product and overheating did not occur during functional testing. The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended. Additional information provided in g3.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVC REPL,MDU, HAND CNTRL, PWRMX
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7151857
MDR Text Key96028895
Report Number1643264-2017-01861
Device Sequence Number1
Product Code HAB
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE
Reporter Occupation HEALTH PROFESSIONAL
Type of Report Initial,Followup
Report Date 02/22/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/28/2017
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number72200616S
Device Catalogue Number72200616S
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer01/13/2018
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/19/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-