Catalog Number 4002C2060
Device Problems Failure to Advance (2524); Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2017
Event Type  malfunction  
Manufacturer Narrative
The product was disposed of by the hospital and is no longer available for return.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a coil embolization procedure in the left internal iliac artery (iia) using penumbra coil 400s (pc400s).It was reported that the patient¿s vessel was slightly calcified and was tortuous.During the procedure, the physician felt a strong touch of "scratching" and was unable to advance a pc400 more than about 30 cm into the px slim delivery microcatheter (px slim); therefore, the px slim was removed with the pc400 inside.The pc400 was removed from within the px slim, and the procedure was then completed using six new pc400s and the same px slim.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key7152749
MDR Text Key96211529
Report Number3005168196-2017-02301
Device Sequence Number1
Product Code HCG
UDI-Device Identifier00814548010762
UDI-Public00814548010762
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120330
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date11/04/2023
Device Catalogue Number4002C2060
Device Lot NumberF66346
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/29/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/06/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-