Results: the neuron select catheter 6f (6f select) was ovalized from 2.0-7.0, 18.0-25.0, 95.0-98.0, and 102.0-105.0 cm from the hub.Conclusions: evaluation of the returned 6f select revealed multiple ovalized regions along the length of the catheter.This damage are likely a result of improper handling during preparation and may have been due to forceful gripping of the catheter during preparation.During functional analysis, a 0.035¿ mandrel, matching the size of the reported guidewire, was unable to advance past the proximal ovalizations confirming the reported complaint.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
While preparing a neuron select catheter 6f (6f select) for a thrombectomy procedure, the physician flushed the 6f select and attempted to load a glidewire through it; however, the glidewire would not advance.Therefore, the 6f select was set aside and the procedure was completed using another 6f select and the same wire.
|