Model Number 3662
Device Problems Premature Discharge of Battery (1057); Device Inoperable (1663)
Patient Problem Inadequate Pain Relief (2388)
Event Date 12/06/2017
Event Type  Injury  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017. The capa was initiated on 2016-02-23 to address the issue of the proclaim ipg entering service application (orion ipg family). Investigation is complete and being monitored by the manufacturer.
 
Event Description
Follow up information revealed patient has not had effective stimulation since spring of 2017. Patient's scs system was replaced with newer model. It was also confirmed that device was in surgery mode prior to surgery. Patient's weight has been added.
 
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported. The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported that patient's ipg was inoperable after lead revision surgery on (b)(6) 2017. As a result, patient's ipg was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROCLAIM 7 ELITE
Type of DeviceSCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7153174
MDR Text Key106791264
Report Number1627487-2017-08784
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date08/11/2019
Device Model Number3662
Device Lot Number6090076
Other Device ID Number05415067020222
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-060217-001-C

Patient Treatment Data
Date Received: 12/28/2017 Patient Sequence Number: 1
-
-