• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE STERLING¿; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE STERLING¿; CATHETER, PERCUTANEOUS Back to Search Results
Model Number H74939031404010
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2017
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4). the device was not returned for analysis.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that balloon rupture occurred.Vascular access was obtained using ipsilateral antegrade approach via the right femoral artery.The target lesion was located in the moderately tortuous and moderately calcified right popliteal artery.After a guide wire crossed the lesion, a 4.0mmx40mmx135cm (4f) sterling¿ balloon catheter was advanced for dilatation.However, during the first inflation at 6 atmospheres, the balloon ruptured.The device was completely removed from the patient and the procedure was completed with a 4.0mmx60mm sterling¿ balloon catheter.No patient complications nor injuries were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING¿
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7154016
MDR Text Key96126468
Report Number2134265-2017-12846
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
PMA/PMN Number
K053118
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 12/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Model NumberH74939031404010
Device Catalogue Number39031-40401
Device Lot Number18619260
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-