Catalog Number H938737 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/04/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that two (2) units of 250ml exactamix eva tpn bags leaked.The event had occurred after the bag had been filled with an unspecified solution.The reporter stated the leak came from where the fill port met the bag.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The supplier of this product has been notified of this condition should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
The device was received for evaluation.Visual inspection, microscopic inspection, and functional testing were performed.Functional testing was performed by filling the bag with water and squeezing the bag; a small leak was observed from the fill port weld.Microscopic inspectional also revealed a leak from the weld.The reported condition was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|