• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - ENGLEWOOD 250 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - ENGLEWOOD 250 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938737
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 12/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 250ml exactamix eva tpn bag had leaked.The event had occurred after the bag had been filled with an unspecified solution.It was reported the leak was located where the fill port met the bag.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Visual inspection with magnification was performed with no issues noted.Functional testing was performed by filling the bag with water and no leaks were observed.The reported issue was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Correction: for the initial mdr is being corrected to (b)(4) 2017.The device was returned and an evaluation is complete.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection with magnification noted that the bag leaked from where the filling port meets the bag.Functional testing was performed by filling the bag with water and a leak was observed from the fill port weld.The reported issue was verified.The cause of the reported issue could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
250 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE - ENGLEWOOD
englewood CO
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7154251
MDR Text Key96048742
Report Number1416980-2017-10061
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477213
UDI-Public(01)00085412477213
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Catalogue NumberH938737
Device Lot Number60091143
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/15/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-