BOSTON SCIENTIFIC - MAPLE GROVE EMERGE¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493919515250
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/13/2017
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr.: the complaint device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that balloon rupture occurred.The 99% stenosed target lesion was located in the moderately tortuous and mildly calcified mid right coronary artery.A 2.50mm x 15mm emerge¿ balloon catheter was advanced for pre-dilatation.However, during the first inflation at 10 atmospheres, the balloon did not inflate.The device was completely removed from the patient and it was noted that the balloon ruptured.The procedure was completed with a different device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE¿
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7154320
MDR Text Key96056941
Report Number2134265-2017-12953
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
PMA/PMN Number
K113220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 12/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Model NumberH7493919515250
Device Catalogue Number39195-1525
Device Lot Number0019942704
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GUIDE CATHETER: HYPERION JR4; GUIDE WIRE: ASAHI SION; INFLATION DEVICE: ENCORE; INTRODUCER SHEATH: TERUMO 7F
-
-