DEPUY ORTHOPAEDICS, INC. 1818910 APG+ STRAIGHT HANDLE INNER; EXTREMITY INSTRUMENTS : SCREWDRIVERS Back to Search Results
Catalog Number 223000024
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/04/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the apg+ handle will no longer hold apg drill guide anymore.No surgical delay.No pieces broken left in patient.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint: (b)(4).Investigation summary: examination of the returned device confirmed the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APG+ STRAIGHT HANDLE INNER
Type of Device
EXTREMITY INSTRUMENTS : SCREWDRIVERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic dr.
warsaw, IN 46582-0988
5743725905
MDR Report Key7154356
MDR Text Key96049920
Report Number1818910-2017-52730
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10603295102977
UDI-Public10603295102977
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number223000024
Device Lot NumberPC967081
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/28/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age62 YR
-
-