Model Number 60-01-00 |
Device Problems
Pumping Stopped (1503); Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/05/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6) a loan drive unit has been provided to the customer.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova (b)(4) received a report that the centrifugal pump system with tubing clamp stopped circulating and an displayed an error message during a procedure.The user switched out the drive unit to continue the case.The replaced drive unit was later tested on a different console and the same error message was displayed.There was no report of patient injury.
|
|
Manufacturer Narrative
|
A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective attachment arm.The technician replaced the part and also retightened the carriage fixing post to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
|
|