RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Event Description
|
During periodic maintenance, the manufacturer¿s international service technician found that the oxygen concentration test revealed that the measured oxygen concentration was out of the specification.There was no patient involvement.
|
|
Manufacturer Narrative
|
During further investigation (fi), a visual inspection of the gas delivery solenoid valve assembly revealed no signs of damage or contamination.The o2 solenoid valve was installed in an fi test ventilator.The high-pressure leak test (pvt test 2), o2 flow accuracy test (pvt test 6) and the o2 accuracy test (pvt test 7) were performed an all passed.The ventilator was run with 60% o2 setting for at least one hour without errors occurring.The determination could not be made that the device failed to meet specifications.The device was not being used for treatment when the reported event occurred, and there is no relationship of the device to the reported problem.The customer returned o2 solenoid valve was installed in an fi test ventilator.The high pressure leak test was executed passed.The oxygen accuracy test and the oxygen accuracy test were executed and both passed.The ventilator was run for two hours without errors occurring.No failure was found.
|
|
Search Alerts/Recalls
|
|
|