Model Number 8637-40
Device Problems Kinked (1339); Aspiration Issue (2883)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/27/2017
Event Type  malfunction  
Manufacturer Narrative
Other relevant components include: product id: 8709sc, serial# (b)(4), implanted: (b)(6) 2011, explanted: (b)(6) 2017, product type catheter.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) via a manufacturer's representative (rep) regarding a patient who was receiving 2000mcg/ml baclofen at 283mcg/day via an implantable infusion pump for an unknown indication for use.It was reported that during a routine pump replacement for elective replacement indicator the doctor was unable to aspirate from the catheter.The catheter was replaced and it was noted that there was a kink in the catheter near the anchor.No symptoms were reported.It was reported the issue was resolved at the time of the report.It was also reported that the patient's status at the time of the report was noted as "alive-no injury." no further complications were anticipated/reported.
 
Manufacturer Narrative
Product id 8709sc lot# (b)(4) implanted: (b)(6) 2011 explanted: (b)(6) 2017 product type catheter.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a manufacturer's representative (rep).The rep reported that the doctor was aware of the issue.The doctor noticed that there was a kink right at the spot near the end of the anchor.No further complications were anticipated/reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7154736
MDR Text Key96124408
Report Number3004209178-2017-26779
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169530126
UDI-Public00643169530126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/14/2019
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/02/2018
Date Device Manufactured08/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age19 YR
-
-