• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIGN FRACTURE CARE INTERNATIONAL SIGN IM NAIL; INTRAMEDULLARY FIXATION ROD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIGN FRACTURE CARE INTERNATIONAL SIGN IM NAIL; INTRAMEDULLARY FIXATION ROD Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Bone Fracture(s) (1870)
Event Date 11/13/2017
Event Type  Injury  
Manufacturer Narrative
A product investigation was performed for this device.The actual device was not returned to the manufacturer for evaluation.The root cause of the non-union is undetermined.The radiographic and clinical data were reviewed by a sign orthopedic surgeon.This failure does not indicate a defect in the product.A minimal risk is associated with this failure.Sign fracture care international continues to monitor these events as part of our post market surveillance activities.
 
Event Description
We became aware on 12/28/2017 that a sign im nail implanted to repair a fracture was replaced due to a non-union.The im nail was replaced with a 10mm x 260mm standard nail per the sign technique manual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGN IM NAIL
Type of Device
INTRAMEDULLARY FIXATION ROD
Manufacturer (Section D)
SIGN FRACTURE CARE INTERNATIONAL
451 hills street
suite b
richland WA 99354 5502
Manufacturer (Section G)
SIGN FRACTURE CARE INTERNATIONAL
451 hills street
suite b
richland WA 99354 5502
Manufacturer Contact
robert schmitt
451 hills street
suite b
richland, WA 99354-5502
5093711107
MDR Report Key7154799
MDR Text Key96050666
Report Number3034525-2017-00149
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeBG
PMA/PMN Number
K022632
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial
Report Date 12/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age27 YR
Patient Weight79
-
-