• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM 5 ELITE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM 5 ELITE; SCS IPG Back to Search Results
Model Number 3660
Device Problem Premature Elective Replacement Indicator (1483)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inaccurate elective replacement indicator advisory notice issued by abbott on 12september2017.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported that diagnostics indicated the elective replacement indicator (eri) triggered earlier than intended.The device has the appropriate level of battery voltage to provide therapy.No intervention planned at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7154856
MDR Text Key96059892
Report Number1627487-2017-08789
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 12/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/29/2018
Device Model Number3660
Device Lot Number5551343
Other Device ID Number05415067020192
Was Device Available for Evaluation? No
Date Manufacturer Received12/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/09/12/2017/001-C
Patient Sequence Number1
Treatment
MODEL: 3228, SCS LEAD
-
-