The ventilator was investigated by our service engineer(fse).Fse could not duplicate the issue during simulated use testing.And no parts were replaced.Our device log evaluation shows that the cause is attributed to the frequent changes of the parameter settings of o2, by the user.The logs show that the user had been changing the parameters up and down, in very short time intervals.The ventilator system takes some time to adapt to these changes, otherwise alarms will be generated for high or low o2 concentrations, as in this case.Our conclusion is that there was no malfunction of the ventilator system at the time of event.(b)(4).Ref.Exemption #: e2018003.(b)(4).
|