The customer stated that she changed the batteries in the coaguchek xs meter serial number (b)(4) six months ago.The meter needed new batteries again, so she looked in the battery compartment.There was corrosion in the battery compartment.The customer removed the batteries and put in one new battery.Upon insertion of the battery, the meter sparked and she noticed burn marks in the battery compartment.She removed the one battery with a pen and left the batteries out of the meter.The customer did not know what type of batteries she had in the meter.The customer did not need any medical treatment and will not seek any medical treatment.No actions were taken based on the event and no adverse events were alleged to have occurred.The customer's meter was requested for investigation and a replacement meter was sent to the customer.
|