• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Model Number XS PST
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2017
Event Type  malfunction  
Manufacturer Narrative
Unique identifier (udi)#: (b)(4).
 
Event Description
The customer stated that she changed the batteries in the coaguchek xs meter serial number (b)(4) six months ago.The meter needed new batteries again, so she looked in the battery compartment.There was corrosion in the battery compartment.The customer removed the batteries and put in one new battery.Upon insertion of the battery, the meter sparked and she noticed burn marks in the battery compartment.She removed the one battery with a pen and left the batteries out of the meter.The customer did not know what type of batteries she had in the meter.The customer did not need any medical treatment and will not seek any medical treatment.No actions were taken based on the event and no adverse events were alleged to have occurred.The customer's meter was requested for investigation and a replacement meter was sent to the customer.
 
Manufacturer Narrative
The meter was returned for investigation.The meter could not be turned on since the battery contacts in the compartment were contaminated due to a battery that had leaked.The device was opened and no abnormalities were observed.The contamination and discoloration in the battery compartment were caused only by leakage of a battery.Marks of increased heat or evidence of fire were not visible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key7156626
MDR Text Key96450376
Report Number1823260-2018-00012
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 02/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberXS PST
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-