MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Failure to Capture (1081); High Capture Threshold (3266)
Patient Problems Dyspnea (1816); Edema (1820)
Event Date 11/30/2017
Event Type  Injury  
Event Description
It was reported that few weeks after device implant, patient was admitted to hospital with congestive heart failure exacerbation.The patient was experiencing weight gain, increasing dyspnea on exertion and ble edema.The left ventricular capture threshold was high.It was noted that the lv lead was placed at inferior posterolateral positon due to patient¿s anatomy.There was left ventricular pacing but it was not enough to compensate.At higher output, the patient would feel simulations at some sites.The lv lead was explanted and an epicardial lead was implanted.The patient condition after the procedure was good.
 
Event Description
New information received notes that loss of capture was also noted.
 
Manufacturer Narrative
Analysis was normal.No anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7158134
MDR Text Key96175477
Report Number2938836-2018-00027
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Model Number1458Q/86
Device Lot NumberA000042981
Other Device ID Number05414734503198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/17/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age85 YR
-
-