(b)(4).Device evaluated by manufacturer: the device was returned for analysis.The device returned is stuck inside of a fg sterling mr device, however, the filterwire has the spring tip damaged (curvy).The outer diameter dimensions were found within specification.Sheathing and unsheathing test was not performed due to the stuck condition.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
|
Same case as mdr id: 2134265-2017-13203.Reportable based on device analysis completed on 13-dec-2017.It was reported that the tip of the device was kinked.The target lesion was located in the moderately tortuous coronary artery.A 190cm filterwire ez¿ was selected for use.During preparation, upon unpacking, the wire tip got kinked while removing the yellow sheath.The procedure was completed with another of the same device.No patient complications reported.However, device analysis revealed that the filterwire was stuck inside an fg sterling mr device.
|