• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. EZ PASS 70 DEGREE LEFT; SUTURE PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. EZ PASS 70 DEGREE LEFT; SUTURE PASSER Back to Search Results
Model Number N/A
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Information (3190)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Not returned to manufacturer.
 
Event Description
The 70 degree ezpass left did not allow the nitinol kite to pass through the end of the passer, therefore making the instrument unusable.The device did not appear to grip to the nitinol wire with enough force to pass the wire through the curved end.Surgery was therefore canceled.No further information has been made available at this time.
 
Manufacturer Narrative
Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Review of the complaint history determined that no further action is required as no were trends identified.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ PASS 70 DEGREE LEFT
Type of Device
SUTURE PASSER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7160946
MDR Text Key96464443
Report Number0001825034-2017-11515
Device Sequence Number1
Product Code HWQ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number904054
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-