• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ANTHEM RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ANTHEM RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3210
Device Problem Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2017
Event Type  malfunction  
Event Description
It was reported that during a normal pulse generator change, both elective replacement indicator and end of service were falsely triggered due to cautery.A firmware was re-downloaded successfully.Due to patient¿s anatomy, the device was replaced the next day.The patient was stable throughout both procedures.
 
Manufacturer Narrative
The reported event of end of service and diagnostic anomaly was not confirmed in the laboratory.The device was tested on the bench under electrical and mechanical conditions and the device exhibited normal characteristics.The reported event is consistent with the device exposure to cautery during the explant procedure.The battery longevity is considered normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHEM RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7172460
MDR Text Key96667961
Report Number2017865-2018-00353
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2015
Device Model NumberPM3210
Device Lot Number4328466
Other Device ID Number05414734503785
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
-
-