Catalog Number UNKNOWN |
Device Problems
Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Internal Organ Perforation (1987); Thrombosis (2100); Blood Loss (2597); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Catalog# is unknown but referred to as cook celect filter.Name and address for importer site: (b)(4).Since catalog# is unknown 510(k) could be either k061815, k073374, k090140, k112119, k121057 or k121629 (b)(4).It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2013".Patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
Exemption number e2016032.(b)(4).Name and address for importer site: (b)(4).Additional information: investigation: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿bleeding, organ perforation, dvt, numbness, anxiety".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Unknown if the reported bleeding, numbness are directly related to the filter.Unknown if the reported anxiety is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Filter celect is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Manufacturer Narrative
|
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
This additional informational received on 02/19/2018: pt allegedly received an implant on (b)(6) 2013 via the right internal jugular vein due to deep vein thrombosis.Pt alleges bleeding, organ perforation (veterbral body).Pt further alleges, post implant deep vein thrombosis, back hemorrhaging, blackouts, body numbness, physical therapy to recover movement, bad walking gate, joints are wearing out prematurely, anxiety.Attempted retrieval on (b)(6) 2017 due to device malfunction.
|
|
Manufacturer Narrative
|
Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Unknown if the reported dvt is directly related to the filter and unable to identify a corresponding failure mode at this time.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: cook medical incorporated (cmi) (b)(4).Registration no.: (b)(4).
|
|
Manufacturer Narrative
|
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Manufacturers ref# (b)(4).Name and address for importer site: cook medical incorporated (cmi), 400 daniels way, bloomington, in 47404, registration no.: 3005580113.Investigation is still in progress.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Additional information received (b)(6) 2019: per ir removal vena cave filter, dated (b)(6) 2017, "the inferior vena cava is widely patent without evidence of caval thrombus.Technically successful retrieval of inferior vena cava filter." per pathology, dated (b)(6) 2017, "received fresh, labeled with the patient's name and designated on the requisition as "ivc filter," is a metallic cone-shaped device measuring 5.0 x 3.5 x 3.0 cm, consistent with ivc filter device.No soft tissue is grossly identified in the specimen.Gross photographs are taken.The specimen is for gross examination only.".
|
|
Search Alerts/Recalls
|
|
|