Model Number DL383400
Device Problem Failure to Shut Off (2939)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer states that the pump will not stop when the feed is complete.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed and the customer states, ¿the pump will not stop when it should.¿ the unit was triaged and the reported issue could not be confirmed at this time.A review of the device history record shows that this unit was manufactured in 2014 and was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
15 hampshire st
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN
15 hampshire st
mansfield MA 02048
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7176145
MDR Text Key96782777
Report Number2183870-2018-00010
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDL383400
Device Catalogue NumberDL383400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/11/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-