PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Device Problem Malposition of Device (2616)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2017
Event Type  Injury  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The product lot number was not provided, therefore, the manufacturing records could not be reviewed.The device was implanted in the patient.
 
Event Description
The patient was undergoing a coil embolization procedure in the internal carotid artery (ica) using penumbra smart coils (smart coils).During the procedure, a smart coil prolapsed from the wide neck aneurysm.The patient was given aspirin, integrin and antiplatelets infusion.The physician then placed an intracranial stent to prevent permanent impairment or damage.The event resolved on the same day, (b)(6) 2017.The wide neck aneurysm coil prolapse requiring stenting was adjudicated to be an adverse event with a probable relationship to the smart coil system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key7176668
MDR Text Key96785713
Report Number3005168196-2018-00094
Device Sequence Number1
Product Code HCG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 12/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
-
-