The customer reported via phone call that they experienced high blood glucose.Customer was able to troubleshoot during the call.The customer¿s blood glucose level was 400, 328, 318, 311, 267, 233, 300, 314 and 205 mg/dl.The customer was treated with an injection.The customer experienced symptoms such as nausea, not fill good.The customer was neither in emergency room, nor admitted into hospital as a result of low blood glucose.Based on customer report customer does allege pump was over delivering.The customer performed high pressure test and it did pass.The customer also states had one little spot about 3" from the reservoir, the spot is about 1/8" long bubble.The insulin pump will not be returned for analysis.Customer reported disconnecting the infusion set at the quick release.Customer reported that insulin exited during the fill tubing process.No product is expected to be returned.
|