Model Number S7 |
Device Problems
No Display/Image (1183); Device Operates Differently Than Expected (2913)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/13/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.It was reported that the monitor of the navigation system was replaced to resolve the reported issue.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect monitor was received by the manufacturer for evaluation.Testing found that the monitor would lose display functionality after being used for an extended period of time.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
|
Event Description
|
A medtronic representative reported that, while outside of a procedure, the monitor of the navigation system would lose functionality after some time.It was noted that the navigation system did not move when the reported issue occurred.There was no patient present when this issue was identified.No additional information was provided.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|