Model Number AB-5100H |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Code Available (3191)
|
Event Date 12/19/2017 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing loss of lock.External equipment was exchanged, however, the issue was not resolved.Revision surgery is scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
The external visual inspection revealed that the array was severed prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed severed electrode wires at the epoxy header region.This is believed to have occurred during revision surgery.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed an electrical test performed.The device failed the residual gas analysis test.The internal visual inspection revealed that a resistor had a corroded trace.It is believed that the failure of this implantable cochlear stimulator (ics) device was caused by a loss of hermetic seal.This is concluded from the residual gas analysis data.Moisture admitted into this device through this leak resulted in the corrosion of the resistive film material and the shift in resistance value of a resistor.This ultimately caused the device to cease functioning.This older device configuration is no longer manufactured.This is the final report.
|
|