• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES INC. OPTION IVC FILTER; RETRIEVABLE INFERIOR VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES INC. OPTION IVC FILTER; RETRIEVABLE INFERIOR VENA CAVA FILTER Back to Search Results
Catalog Number 352506070E
Device Problems Material Fragmentation (1261); Device Or Device Fragments Location Unknown (2590)
Patient Problem No Information (3190)
Event Date 12/11/2017
Event Type  malfunction  
Manufacturer Narrative
The lot number was not provided, so the device history record could not be reviewed.Photos of the retrieved object were provided by the customer; however, the actual product was not returned for evaluation.It appears that the introducer sheath lining may have delaminated but that cannot be confirmed without inspecting the actual product.Since the device is not available for return, no further evaluation can be conducted at this time.
 
Event Description
During jugular filter placement, an object came out of the introducer sheath.The piece was retrieved; however, the patient is still in the hospital and being monitored because it is unknown if all pieces were retrieved.
 
Manufacturer Narrative
The sheath lining filament was returned for review; however, the filter, sheath, or pusher were not returned.Without the filter and/or entire sheath assembly available for evaluation, further conclusions cannot be drawn.Since the main device is not available for return, no further evaluation can be conducted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTION IVC FILTER
Type of Device
RETRIEVABLE INFERIOR VENA CAVA FILTER
Manufacturer (Section D)
ARGON MEDICAL DEVICES INC.
1445 flat creek road
athens TX 75751
Manufacturer (Section G)
ARGON MEDICAL DEVICES INC.
Manufacturer Contact
gail smith
1445 flat creek road
athens, TX 75751
2144368995
MDR Report Key7180525
MDR Text Key96961923
Report Number1625425-2017-00190
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133243
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number352506070E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-