DEPUY SYNTHES PRODUCTS LLC OSCILLATING SAW ATTACHMENT II; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
Catalog Number 511.801 |
Device Problems
Device Operates Differently Than Expected (2913); Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/08/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the oscillating saw attachment device stopped working and noise was coming from the device when the trigger was pressed.It was further reported that the battery device was changed; however, the issue still persisted.It was unknown if there was a delay to the surgical procedure.The procedure was completed with a spare device.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.It was determined that the coupling tool side was not functioning and defective.It was further determined that the tool holder was defective.It was further determined that the device failed pretest for check the machine coupling and check symmetry with handpiece.The assignable root cause was determined to be due to improper maintenance, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|