MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAS
Device Problem Fluid/Blood Leak (1250)
Patient Problem Hyperglycemia (1905)
Event Date 01/02/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump alarmed unexpected restart.The customer reported that when they replace the battery it will come up with a blank screen and it will say 12 am.This time it was coming up the battery full, the insulin was empty and it's coming up unexpected restart.Customer stated this was the fourth time she had an issue when she changed the battery.Customer stated she does not know what her blood glucose was at the time.Customer stated she got home and could smell insulin leaking.Customer stated she checked her blood glucose at the time and it was 250mg/dl.Customer stated noticed the insulin leaking at her infusion site.Customer states a temp basal was not programmed before the unexpected restart alarm occurred.Alarm occurred shortly after inserting a new battery.The customer declined to troubleshoot for high blood glucose.Customer bolused with her pump to treat for high blood glucose.The customer was advised that the insulin pump needed to be replaced.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Pump passed the displacement, self test and no unexpected restart error alarm noted during test.Pump received with cracked case at the display window corner, broken reservoir tube lip, minor scratched lcd window, stained end cap sticker, stained address/serial number label, cracked belt clip slot, cracked case reservoir tube window corner, cracked lcd window and cracked battery tube threads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAS
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7186970
MDR Text Key97668817
Report Number3004209178-2018-43061
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169404342
UDI-Public(01)00643169404342
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751NAS
Device Catalogue NumberMMT-751NAS
Device Lot NumberA2751NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received04/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age64 YR
Patient Weight195
-
-