• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED INFST MM T-975T MIO 2PK 9MM 32IN CLEAR; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED INFST MM T-975T MIO 2PK 9MM 32IN CLEAR; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number MMT-975T
Device Problem Bent (1059)
Patient Problem Hyperglycemia (1905)
Event Date 03/24/2016
Event Type  Injury  
Manufacturer Narrative
Note: this is a remediation mdr.Medtronic diabetes implemented revised mdr reportability criteria effective on october 1, 2017.Subsequently, medtronic diabetes conducted a two year retrospective review of complaints.This event was retrospectively identified to be reportable based on the revised mdr reportability criteria.Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
Customer reported via phone call that they had a high blood glucose level of over 400 mg/dl.Customer reported that they had a bent infusion set cannula.This occurred on the first day of use.Customer treated the high blood glucose with an injection of insulin.No product is expected to be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFST MM T-975T MIO 2PK 9MM 32IN CLEAR
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire st.
northridge, CA 91325-1219
MDR Report Key7187804
MDR Text Key97267831
Report Number2032227-2018-00644
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 01/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-975T
Device Catalogue NumberMMT-975T
Device Lot Number5131870
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/24/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age55 YR
-
-