Model Number FUSION
Device Problem No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2017
Event Type  Injury  
Manufacturer Narrative
After additional review of this event it has been identified that this event meets the definition of a serious injury. If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment. The computer of the navigation system was reported to have been replaced. The hardware, software, and instruments passed the system checkout. The system was found to be fully functional. The computer of the navigation system was returned to the manufacturer for analysis. The device was found to be fully functional with no problem found. The reported event could not be duplicated by medtronic personnel.
 
Event Description
A medtronic representative reported that, while in a functional endoscopic sinus surgery (fess), the navigation system displayed that there was no input detected on the monitor. It was reported that the issue occurred when starting the system. The navigation system was plugged in on one side of the operating room (or) where it operated as designed, after switching outlets, the reported issue occurred. The surgeon opted to complete the procedure without the use of the navigation system. There was a reported delay to the procedure of less than 1 hour due to this issue. There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION NAVIGATION SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7188076
MDR Text Key141498903
Report Number1723170-2018-00219
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberFUSION
Device Catalogue Number9733560
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/07/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 01/12/2018 Patient Sequence Number: 1
-
-