• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND ACCOLADE IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND ACCOLADE IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problems Over-Sensing (1438); Pacing Problem (1439); Ambient Noise Problem (2877)
Patient Problems Syncope (1610); Bruise/Contusion (1754); Fall (1848); Dizziness (2194)
Event Date 09/04/2017
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy pacemaker (crt-p) and another manufacturer's right ventricular (rv) lead exhibited noise and oversensing that resulted in pacing inhibition for greater than 2 seconds of asystole. The patient experienced dizziness and syncope and suffered a fall resulting in bruising to the head, shoulder and arm on the right side of the body and resulted in three days of hospitalization. The device was noted to be programmed vdir. Boston scientific technical services (ts) reviewed the episodes of noise and discussed that the noise was consistent with oversensing related to the respiratory rate trend (rrt) feature. Ts discussed troubleshooting options; including programming rrt off. It was reported that the device was reprogrammed to voo and rrt was programmed off. No additional adverse patient effects were reported. This product remains implanted and in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACCOLADE
Type of DeviceIMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7189565
MDR Text Key97278256
Report Number2124215-2017-20721
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial
Report Date 10/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date11/03/2018
Device Model NumberU128
Other Device ID NumberVALITUDE X4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received10/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/03/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 01/15/2018 Patient Sequence Number: 1
-
-