Catalog Number 328821 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/03/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medical device expiration date: unknown.No lot # provided.Initial reporter phone#: (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
|
Event Description
|
It was reported that during use, a bd ultra-fine¿ insulin syringe malfunctioned as the needle separated from the syringe.There was no report of injury or medical intervention needed.
|
|
Manufacturer Narrative
|
Investigation summary: customer returned a photo of a 1cc syringe.Customer states that the catheter split.The photo was examined and exhibited the cannula separated from the syringe and stuck in the vial.Capa (b)(4) has been opened to address this issue for 1ml syringes.Unable to perform complaint lot history check due to unknown lot number.Investigation based on the samples / photo(s) received the investigation concluded: confirmed: bd was able to duplicate or confirm the customer¿s indicated failure (cannula separates).Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: probable root cause determined to be misalignment during application of adhesive on the needle lines.When this occurs, adhesive runover onto the hub or possible the cannula may occur.Additionally, adhesive may be inaccurately applied to the rubberized pull wheel on the line, which can additionally transfer adhesive to the cannula during routine use.Capa (b)(4) has been opened to address this issue for 1ml syringes.
|
|
Search Alerts/Recalls
|