BD MEDICAL - DIABETES CARE BD ULTRA-FINE¿ INSULIN SYRINGE Back to Search Results
Catalog Number 328821
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2017
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.No lot # provided.Initial reporter phone#: (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event Description
It was reported that during use, a bd ultra-fine¿ insulin syringe malfunctioned as the needle separated from the syringe.There was no report of injury or medical intervention needed.
 
Manufacturer Narrative
Investigation summary: customer returned a photo of a 1cc syringe.Customer states that the catheter split.The photo was examined and exhibited the cannula separated from the syringe and stuck in the vial.Capa (b)(4) has been opened to address this issue for 1ml syringes.Unable to perform complaint lot history check due to unknown lot number.Investigation based on the samples / photo(s) received the investigation concluded: confirmed: bd was able to duplicate or confirm the customer¿s indicated failure (cannula separates).Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: probable root cause determined to be misalignment during application of adhesive on the needle lines.When this occurs, adhesive runover onto the hub or possible the cannula may occur.Additionally, adhesive may be inaccurately applied to the rubberized pull wheel on the line, which can additionally transfer adhesive to the cannula during routine use.Capa (b)(4) has been opened to address this issue for 1ml syringes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ULTRA-FINE¿ INSULIN SYRINGE
Type of Device
INSULIN SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7190559
MDR Text Key97882482
Report Number1920898-2017-00449
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K024112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,other
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 03/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number328821
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? No
Date Manufacturer Received12/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-