Model Number S7
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/18/2017
Event Type  malfunction  
Manufacturer Narrative
Patient age, weight and sex were requested, but not provided. A medtronic representative performed a system checkout. He was unable to recreate the problem. He opened all the panels, moved the system around, and jiggled all the cables to check for loose connections, but nothing caused the monitors to go black. It seems to have been a one time issue as the site has been using the system since and it has been working fine. Unable to replicate reported issue. No parts replaced.
 
Event Description
A nurse reported that during a cranial resection, the navigation system became unresponsive in the navigate screen, and the monitors went black. Both monitors went out at the same time. The issue was brief and the monitors came back so the site was able to continue the case. No harm to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S7 SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
catherine eaton
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902092
MDR Report Key7196039
MDR Text Key97668965
Report Number1723170-2018-00245
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation
Type of Report Initial
Report Date 01/16/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/16/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-