The following controls are in-place to mitigate ¿broken blade¿ condition at aspen surgical las piedras site: ¿ heat treatment in-process at the beginning and end of each lot are inspected for dimension integrity using a pin gauge, flatness gauge and perforation length gauge, ductility test, and hardness test.
¿ heat treatment quality inspections at the beginning and end of each lot are inspected for dimension integrity using a pin gauge, flatness gauge and perforation length gauge, ductility test, and hardness test.
¿ 100% visual inspection by certified personnel is performed to segregate nonconforming material, including broken blades, prior to assembly and packaging process.
Each associate performing 100% inspection process is certified and re-certified on an annual basis on blade defect criteria¿s.
¿ capa (b)(4) was initiated to evaluate current controls in place corrective actions identified were completed and capa closed on 10/26/2015.
¿ capa (b)(4) was also initiated to continue in the monitoring and improvement of current control.
Device not returned.
|