LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/28/2015
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).This event was initially considered to be non-reportable.However, after additional evaluation, livanova (b)(4) has decided to reclassify this event as reportable.This report is being filed as part of a retrospective review conducted in response to this new decision.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).After the procedure the heater-cooler system 3t was tested in the workshop of technical department, where the fuses were blown.A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced it back to the compressor.Signs of corrosion were noticed on the cooling coils.Corrective actions are in progress for this type of issue.The unit was recommended to be returned to livanova (b)(4) for further investigation and repair.After a follow-up communication, livanova was informed, that the customer replaced the claimed device with one from another manufacturer and that the device was scrapped.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.Device evaluated on site by livanova technician.
 
Event Description
Livanova (b)(4) received a report that the heater cooler system 3t displayed high leakage in the or during a procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7202970
MDR Text Key98071795
Report Number9611109-2018-00049
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeNL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 01/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/29/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/23/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-