WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-1-JUG-CELECT-PT
Device Problems Bent (1059); Unintended Movement (3026); Insufficient Information (3190)
Patient Problems Vessel Or Plaque, Device Embedded In (1204); Pain (1994); Thrombus (2101); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(4).Catalog# is unknown but referred to as cook celect filter.(b)(6).Since catalog# is unknown 510(k) could be either k061815, k073374, k090140, k112119, k121057 or k121629.It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2016".Patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
Exemption number e2016032.(b)(4).Corrected data: adverse event and product problem to product problem.Outcomes to adverse event: other to .It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'celect-pt, tilt, unable to retrieve, pain, permanent/ continuous injuries, blood clot, bent strut'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Ivc thrombotic occlusion as an outcome for cook ivc filters is addressed in the published scientific literature.Ivc thrombotic occlusion is defined as the presence of an occluding thrombus in the ivc after filter insertion and documented by ultrasound (us), ct, mr imaging or venography; this may be symptomatic or asymptomatic.Unknown if the reported permanent/ continuous injuries is directly related to the filter and unable to identify corresponding failure mode(s) at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable, or unchanged.Exemption number e2016032.(b)(4).510 (k) k121629 this report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
This additional information received on 16jan2018 as follows: ¿patient [pt] allegedly received an implant on (b)(6) 2016 via the right internal jugular vein due to deep vein thrombosis (left leg),recent postoperative hematoma.[pt] is alleging tilt and pain, permanent and continuous injuries without further explanation.[pt] further alleges blood clot without further explanation.Filter retrieval was attempted on (b)(6) 2016.The retrieval attempt was unsuccessful.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT® PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key7205017
MDR Text Key97712035
Report Number3002808486-2018-00147
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K121629
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup
Report Date 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-1-JUG-CELECT-PT
Was Device Available for Evaluation? No
Distributor Facility Aware Date01/09/2018
Date Manufacturer Received02/07/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
Patient Weight67
-
-