Same case as mdr id: 2134265-2017-13373 and 2134265-2017-13374.
It was reported that foreign matter was observed on the rotawire and a rotawire fracture occurred.
The target lesion was located in the anterior tibial artery of the right lower leg.
A peripheral rotawire and a peripheral rotalink plus were selected for use.
There were two rotawires that were pulled out from its hoop and white debris was observed at the proximal end.
Subsequently, the second rotawire was used during platforming.
When the physician loaded the rotaburr onto the rotawire, it was noted that the wire snapped in half.
The rotawire was exchanged with another of the same device and the rotaburr was loaded onto it.
However, the burr was unable to load onto the wire.
The procedure was completed with another of the same rotaburr.
No patient complications were reported and the patient's status was fine.
|