Intuitive surgical, inc.
(isi) received the instrument involved with this complaint and completed the device evaluation.
Failure analysis confirmed the customer reported failure.
The instrument was found to have a broken pitch cable at the distal clevis hub.
The broken cable segment that contains the crimp was missing from clevis approximately 0.
032 x 0.
046 in size.
The clevis did not exhibit any damage or wear marks.
Other cables at the wrist were not damaged.
Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event.
The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis suggests that if the malfunction were to recur it could cause or contribute to an adverse event.
|