Catalog Number 1012454-15
Device Problems Detachment Of Device Component (1104); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/08/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported complaints appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that an nc trek was advanced to the circumflex coronary artery lesion and inflation was performed without issue.The balloon was fully deflated prior to removal.During device removal through the 6french guide catheter, resistance was met with the guide catheter and the balloon separated.All was removed as a single unit without any intervention.There were no adverse patient effects and there was no clinically significant delay.There was no additional information provided regarding this issue.
 
Manufacturer Narrative
Internal file number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7209686
MDR Text Key97957550
Report Number2024168-2018-00499
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648152115
UDI-Public08717648152115
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number1012454-15
Device Lot Number70720G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/25/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial