• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC HUDSON REDUCTION DRIVE UNIT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC HUDSON REDUCTION DRIVE UNIT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.018
Device Problems Device Inoperable (1663); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/29/2017
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.The actual device has been returned and is currently pending evaluation.Once the evaluation of the device has been completed, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported that during an unspecified surgical procedure of the hip it was observed that the reduction drive device would not go in reverse.It was reported that there was no delay in the procedure due to the event was an unspecified spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.The actual device was returned for evaluation.The device was evaluated and the reported condition was not confirmed.An assessment was performed and the device met all manufacture's specifications.Therefore, an assignable root cause was not determined.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON REDUCTION DRIVE UNIT
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 33410
6107195000
MDR Report Key7212965
MDR Text Key98190991
Report Number8030965-2018-50567
Device Sequence Number1
Product Code HWE
UDI-Device Identifier10886982240521
UDI-Public(01)10886982240521(11)140625(10)MA1049
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.018
Device Lot NumberMA1049
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received02/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-