Catalog Number B2070-200
Device Problems Crack (1135); Inflation Problem (1310); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/03/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).Evaluation summary: visual and functional inspection was performed on the returned device.The reported crack was not confirmed; however, the inflation issue was confirmed due to a leak in the hub.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.Av conducted root cause analysis and determined the issue may be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Av will continue to trend the performance of these devices.
 
Event Description
It was reported that the procedure was to treat an occluded previously deployed stent in the superficial femoral artery (sfa).The 7x200 armada 35 balloon dilatation catheter (bdc) was advanced to the lesion with no resistance.The bdc was not prepped prior to insertion.An attempt was made to inflate the balloon however it would not inflate due to a crack in the hub.The device was replaced with a new balloon catheter of the same size to complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7213965
MDR Text Key98211663
Report Number2024168-2018-00544
Device Sequence Number1
Product Code LIT
UDI-Device Identifier08717648155413
UDI-Public08717648155413
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 01/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Catalogue NumberB2070-200
Device Lot Number41002G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-