MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problems Device Stops Intermittently (1599); Communication or Transmission Problem (2896); Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/03/2018
Event Type  malfunction  
Manufacturer Narrative
Product, unique device identification (udi), device manufacture date and associated fields updated to proper value.
 
Manufacturer Narrative
The fda codes were updated to reflect current coding guidelines. If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment. The representative was unable to replicate the reported issue. The hardware, software, and instruments passed the system checkout. The system was found to be fully functional.
 
Event Description
A site representative reported that, while in a functional endoscopic sinus surgery (fess), the emitter for the navigation system would intermittently not track. It was reported that the site restarted the navigation system to restore functionality. It was reported that the emitter turned off without prompt from the user and functionality came back without prompt from the user after roughly fifteen seconds. There was no impact on patient outcome. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION NAVIGATION SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7217239
MDR Text Key103085195
Report Number1723170-2018-00381
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup,Followup
Report Date 06/06/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/21/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-