Results: there was no visible damage to the artemis neuro evacuation device (artemis) unit.There was some clotted blood present inside the aspiration chamber.Conclusions: evaluation of the returned artemis device revealed it was able to perform in all aspects within specification.When activating the rotational wire the noise varied noticeably.The change in noise did not affect the function of the device.The root cause of the changes in sound produced by the handle could not be determined.Penumbra artemis devices are inspected at incoming quality control which includes a visual inspection as well as a verification of test results to ensure specifications for each output are met.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a microneurosurgery procedure using an artemis neuro evacuation device (artemis).During the procedure, the physician noticed that there was a decrease in aspiration power and decreased audible and tactile feedback after about 15 minutes of use.The physician therefore flushed the cannula with saline and cleared the aspiration tubing, however no improvement was noticed.The procedure was therefore completed using a new artemis.There was no report of an adverse effect to the patient.
|