MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751LNAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751LNAL
Device Problem Inaccurate Delivery (2339)
Patient Problem Hyperglycemia (1905)
Event Date 01/18/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they experienced high blood glucose.The customer¿s blood glucose level was 463 mg/dl.The customer also claims that every time he tried to deliver a bolus it stops at around 1 to 2 units.The customer declined troubleshoot for high blood glucose.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Insulin pump passed the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test.No excessive no delivery alarm and no motor error alarm noted during test.No bolus stopped alarm noted.The information provided in the section of concomitant product with the initial report was incorrect.The correct information has been included with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751LNAL
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7224598
MDR Text Key98443109
Report Number3004209178-2018-46662
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169507760
UDI-Public(01)00643169507760
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 08/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-751LNAL
Device Catalogue NumberMMT-751LNAL
Device Lot NumberA4751LNALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received07/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age43 YR
Patient Weight179
-
-