• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED Back to Search Results
Catalog Number 532.001
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the small battery drive device did not work even after changing the battery.It was not reported if there were any delays in the surgical procedure or whether a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Corrected data: the date the device was received by the manufacturer was reported as january 12, 2018 in the initial report and has been updated to february 1, 2018.Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device motor was seized, jammed and heavy moving.Therefore the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7225012
MDR Text Key98641038
Report Number8030965-2018-50649
Device Sequence Number1
Product Code HWE
UDI-Device Identifier7611819107083
UDI-Public(01)7611819107083(11)110621
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2018
Date Manufacturer Received03/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-