SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616SR
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/23/2018
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation. There was a relationship found between the returned device and the reported incident. A visual inspection was performed and no issues were noted. Product failed functional testing with hand piece error when inserted into mdu receptacle on test control unit. Cause of hand piece errors is a stuck oscillate button assembly. Defective button switch is making the button assembly stick in the ¿on¿ position. Button switch has a small burr inside causing the keeper switch to stick when button assembly is fully pressed to activate motor and direction. The complaint was confirmed and the root cause has been determined to be a defective button switch. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Event Description
It was reported that the oscillate button was stuck in. No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVCE REPL MDU CNTRL PWRMX ELITE
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7227683
MDR Text Key130368484
Report Number1643264-2018-00081
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/09/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number72200616SR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/08/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-