Model Number S7
Device Problem Material Frayed (1262)
Patient Problem No Patient Involvement (2645)
Event Date 01/10/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.During the onsite inspection, the medtronic representative reported that the metal from the cable was showing and frayed.The wires were not completely insulated.The cable was replaced.The replaced part has not been received by the manufacturer for further analysis.A successful system checkout was performed which verified that the issue had been resolved.
 
Event Description
A medtronic representative reported that the power cable was fraying.There was no damage to the ports observed.No patient was present during the time of the reported incident.
 
Manufacturer Narrative
Additional information: device manufacture date provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902447
MDR Report Key7232580
MDR Text Key98740102
Report Number1723170-2018-00489
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Equipment Company Technician/Representative
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? No
Date Manufacturer Received02/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/20/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-