Model Number MT20649
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/04/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The receiver was returned for evaluation. An exterior visual inspection was performed and failed due to usb connector damage. The receiver failed to charge and boot. The receiver log failed to download due to the damaged usb connector damage. The reported event of the overheating of the receiver was not confirmed. A probable cause could not be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018, the receiver was overheating. No additional event or patient information is available. The receiver has been received for evaluation. A follow up report will be submitted when the evaluation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7233131
MDR Text Key117891783
Report Number3004753838-2018-011185
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/04/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMT20649
Device Catalogue NumberSTK-GL-001
Device Lot Number5091232
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/23/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-