Model Number V200
Device Problem Occlusion Within Device (1423)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.
 
Event Description
The customer reported the ventilator alarmed due to an occlusion-svo.No patient involvement reported.
 
Manufacturer Narrative
The field service engineer (fse) confirmed the reported problem.The fse opened the machine, cleaned the patient circuit and the issue resolved.The determination could not be made that the device failed to meet specifications.The device was not being used for treatment when the reported event occurred, and there is not a relationship of the device to the reported problem.Due to no part returning for failure investigation the root cause of the reported issue could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V200 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
robert corning
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key7234120
MDR Text Key98869763
Report Number2031642-2018-00200
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00884838010970
UDI-Public(01)00884838010970
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-